| ABSTRACT |
|---|
Methods: A total of 102 health care professionals working in the critical care environment were asked to estimate and record the pupil size of a series of 12 artificial eyes with varying pupil diameter and iris colour. All estimations were performed indoors under ambient lighting conditions.
Results: Our results established a statistically significant difference between staff groups in the estimation of pupil size.
Conclusion: The demonstrated variability in pupil size estimation may not be clinically significant. However, it remains desirable to have consistency of measurement throughout the patient care pathway.
Keywords: estimation; pupil measurement; variability
Serial assessment of pupil size and response to light is important during the initial examination and subsequent monitoring of the head-injured patient and is emphasised in all guidelines relating to emergency care.1?3
As the patient progresses from initial injury through acute care to discharge, pupil assessment is performed by different healthcare professionals in many different situations. Current practice is to estimate pupil size manually and studies have shown that there is an element of subjectivity in the values recorded.4,5 Whilst this may not be significant with a single observer, it could introduce error within a multiple observer series. Given that the detection of a changing pupil state rather than its absolute size may require immediate intervention, it is important that measurements are reliable and inter-operator variability low. The primary objective of this study was to compare measurements of pupil size by a variety of different staff groups working throughout the critical care pathway to determine whether there was any statistically significant difference between the groups that could impact on patient care.
| METHODS |
|---|
A total of 102 members of staff from Newcastle Hospitals NHS Trust and the North East Ambulance Service were asked to examine the artificial eyes indoors under ambient light and record their estimations of pupil size. The eyes were presented in random order with a variation in size and colour between each consecutive eye.
The staff members were recruited from accident and emergency departments, intensive care departments, neurosurgical wards, and the ambulance service. There were a total of 68 nurses, 13 qualified medical staff, and 21 ambulance staff. The hospital based personnel consisted of 25 staff from accident and emergency departments, 12 from neurosurgical wards, 22 from neurosurgical intensive care, and 22 from general intensive care.
Univariate analysis of variance was used to compare observations between staff groups while taking into account the artificial eye being observed. Post-hoc comparisons were undertaken using Neuman-Keuls procedure. All analyses were conducted using SPSS version 11.0.
| RESULTS |
|---|
|
| CONCLUSION |
|---|
In the study, artificial eyes have been used to permit a controlled measurement environment, but this necessarily imposes certain limitations. In human subjects, factors such as movement, anatomy, and environment would introduce further random measurement error. As pupil assessment forms an essential part of the neurological examination, ideally it should be both accurate and repeatable even when undertaken by a wide range of healthcare professionals in many different settings.
There are several important physiological measurements (for example, blood pressure and oxygen saturation) that are routinely taken using automatic rather than manual processes. This has greatly facilitated their use and adoption in many different clinical areas and, to some extent, standardised both the measurement method and precision. A device that automatically measures pupil size could not only reduce the demonstrated potential differences in this basic neurological examination, but also provide other important values related to pupil dynamics.4,6